UF V (E3) 10/2020 Page 29/119
To move the freezer through narrow passages (doors, narrow corridors), open the chamber door:
Figure 14: UF V with open chamber door
For transport outside a building use technical equipment (chap. 3.2.2).
3.2.2 Transport outside a building
Before moving the chamber unlock the front castors. The castors are designed only for moving the chamber
inside a building (respect the information given in chap. 3.2.1).
After operation, please observe the guidelines for temporarily decommissioning the chamber (chap. 26.2).
CAUTION
Risk of injury and damages by sliding or tilting of the chamber due to improper
transportation.
Injuries, damage to the chamber.
Do NOT lift or transport the chamber using the door, the door lock and controller hous-
ing or the lower housing.
Do NOT lift the chamber by hand
If possible, avoid transporting the chamber horizontally. It may be transported lying
down ONLY on the hinge side or on its back, but must then stand upright for at least
24 hours before turning on.
Transport the chamber only in its original packaging.
Secure the chamber with transport straps for transport.
Place the shelves on top of each other on the bottom of the interior.
Lift the chamber using technical devices (fork lifter) and place it on the transport pallet.
Set the fork lifter laterally or from the rear in the middle of the chamber. Make sure to
place all the lateral supports of the chamber on the forks (check: the fork protrudes at
the opposite chamber side).
Transport chambers ONLY with the original transport pallet. Set the fork lifter only to
the pallet. Without the pallet the chamber is in imminent danger of overturning
Wear suitable shoes (safety shoes).
Permissible ambient temperature range for transport: -10 °C / 14°F to +60 °C / 140°F.
You can order transport packing and rolling pallets for transportation purposes from BINDER Service.
830 mm
900 mm
1400 mm
1000 mm (UF V 500)
1280 mm (UF V 700)
Loading ...
Loading ...
Loading ...