Binder UF V 700 UF V 500-liter Ultra Low Temperature Freezer

User Manual - Page 30

For UF V 700.

PDF File Manual, 119 pages, Read Online | Download pdf file

UF V 700 photo
Loading ...
Loading ...
Loading ...
UF V (E3) 10/2020 Page 30/119
Wear suitable shoes (safety shoes) during transport.
Following transport, wait at least 8 hours until start-up.
3.3 Storage
Intermediate storage of the chamber is possible in a closed and dry room. Observe the guidelines for tem-
porary decommissioning (chap. 26.2).
Permissible ambient temperature range for storage: -10 °C / 14°F to +60 °C / 140°F.
Permissible ambient humidity: max. 70% r.h., non-condensing
Secure the chamber against unintentional rolling by locking the front castors.
The freezer must stand upright in order to avoid oil running out of the engine casing and resulting dam-
ages to the cooling system. Max. angle of inclination: 10°.
When after storage in a cold location you transfer the chamber to its warmer installation site, condensation
may form in the inner chamber or on the housing. Before start-up, wait at least one hour until the freezer
has attained ambient temperature and is completely dry. According to the type of transport that has taken
place (chap. 3.2) you may have to wait at least 8 hours until start up.
3.4 Location of installation and ambient conditions
The freezer is designed for setting up inside a building (indoor use). Set up the chamber on a flat, even
surface, free from vibration and in a well-ventilated, dry location. Lock the front castors and align the cham-
ber using a spirit level. The site of installation must be capable of supporting the chamber’s weight (see
technical data, chap. 27.3).
NOTICE
Danger of overheating due to lack of ventilation.
Damage to the chamber.
Do NOT install the chamber in unventilated recesses.
Ensure sufficient ventilation for dispersal of the heat.
Observe the prescribed minimum distances when installing the chamber.
NOTICE
Danger to the environment by leakage of refrigerant in the event of a chamber de-
fect.
Alteration of the environment.
Ensure sufficient ventilation of the installation site.
Permissible ambient temperature range for operation: +18 °C / 64.4 °F to +32 °C / 89.6 °F. At elevated
ambient temperature values, fluctuations in temperature can occur.
The ambient temperature should not be substantially higher than the indicated ambient tem-
perature of +22 +/- 3 °C / 71.6 °F +/- 5.4 °F to which the specified technical data relate. For
other ambient conditions, deviations from the indicated data are possible.
Prevent the freezer from sucking warm air from other devices.