Specifications
Safety
8
127
Safety
Designed for 1000 V Measurement Category III, 600 V
Measurement Category IV (with supplied 10:1 probes) per:
EN/IEC 61010-1, Pollution Degree 2
EN/IEC61010-2-030
IEC61010-031
Max. Input Voltages
BNC Input A, B, (C, D) directly ................... 300 V CAT IV
Via VPS410 .............................................. 1000 V CAT III
600 V CAT IV
METER/EXT banana input ....................... 1000 V CAT III
600 V CAT IV
Max. Floating Voltage
FLUKE 190-xxx (test tool or test tool + VPS410)
From any terminal to earth ground ......... 1000 V CAT III
600 V CAT IV
Between any terminal ............................. 1000 V CAT III
600 V CAT IV
FLUKE 190-xxx + VPS510
From any terminal to earth ground ........... 300 V CAT III
Between any terminal ............................... 300 V CAT III
Voltage ratings are given as “working voltage”. They
should be read as Vac-rms (50-60 Hz) for AC sinewave
applications and as Vdc for DC applications.