Specifications
Automatic Scope Measurements
8
119
V/Hz
Purpose .................... to show the measured Vpwm value
(see Vpwm) divided by the fundamental frequency
on Variable AC Motor Speed drives.
Accuracy ................................................... %Vrms + %Hz
Note
AC motors are designed for use with a rotating
magnetic field of constant strength. This strength
depends on the applied vottage (Vpwm) divided
by the fundamental frequency of the appied
voltage (Hz). The nominal Volt and Hz value are
shown on the motor type plate.
Power (A and B, C and D)
Power Factor ........................ ratio between Watts and VA
Range ........................................................ 0.00 to 1.00
Watt ................................... RMS reading of multiplication
corresponding samples of input A or C (volts)
and Input B or D (amperes)
Full Scale Reading ....................................... 999 counts
VA ................................................................ Vrms x Arms
Full Scale Reading ....................................... 999 counts
VA Reactive (VAR) .......................................
((VA)
2
-W
2
)
Full Scale Reading ....................................... 999 counts
Phase (A and B, C and D)
Range ............................................. -180 to +180 degrees
Resolution .......................................................... 1 degree
Accuracy
0.1 Hz to 1 MHz ........................................... ±2 degrees
1 MHz to 10 MHz ......................................... ±3 degrees
Temperature (TEMP)
With Optional Temperature Probe (
°
F not for Japan)
Ranges (°C or °F) .................................. -40.0 to +100.0 °
-100 to +250 °
-100 to +500 °
-100 to +1000 °
-100 to + 2500 °
Probe Sensitivity ............................. 1 mV/°C and 1 mV/°F
Accuracy ........................................... ±(1.5 % + 5 counts)
(add temperature probe accuracy for overall accuracy)
Decibel (dB)
dBV ................................................ dB relative to one volt
dBm .................... dB relative to one mW in 50 or 600
dB on .......................................... VDC, VAC, or VAC+DC
Accuracy .......................... same as VDC, VAC, VAC+DC