70
a. the current channel may be switched in the channel switching menu;
b. the current channel menu can be moved anywhere on the screen;
c. only the open channel is displayed in the channel switching menu;
d. when the math or reference waveform is opened, the current channel switching menu is automatically opened.
5.2 Adjust Vertical Sensitivity
Tap the vertical sensitivity or buttons on the right side of the channel icon to adjust the vertical display of
the waveform corresponding to the channel, so that the waveform is displayed on the screen at an appropriate size.
The vertical sensitivity scale (V/div) after each adjustment is displayed on the channel icon. For example,
means that the current vertical sensitivity of CH1 is 1.0V/div.
The vertical sensitivity coefficient adjusts the vertical sensitivity of the analog channel in steps of 1-2-5 (the probe
attenuation coefficient is 1X), and the vertical sensitivity range of 1:1 probe is 1mV/div-10V/div (optionally
minimum at 500uV/div).