Micsig STO1152C+ STO1152C Plus digital oscilloscope

Micsig STO series user manual - Page 41

For STO1152C+.

PDF File Manual, 353 pages, Read Online | Download pdf file

STO1152C+ photo
Loading ...
Loading ...
Loading ...
26
Source can be automatically triggered, and the triggered source channel can be automatically set to select priority to
the current signal or to the maximum signal.
Open the main menu. Tap “Auto” to open the auto set menu, including channel open/close setting, threshold voltage
setting and trigger source setting.
Figure 2-9 Open Auto Set
Automatic configuration includes: single channel and multiple channels; automatic adjustment of the horizontal
time base, vertical sensitivity and trigger level of signal; the oscilloscope waveform is inverted off, the bandwidth
limit sets to full bandwidth, it sets as DC coupling mode, the sampling mode is normal; the trigger type is set to
edge trigger and the trigger mode is automatic.