Micsig STO1152C+ STO1152C Plus digital oscilloscope

Micsig STO series user manual - Page 343

For STO1152C+.

PDF File Manual, 353 pages, Read Online | Download pdf file

STO1152C+ photo
Loading ...
Loading ...
Loading ...
328
7. If the display becomes slower after setting average times of sampling:
If the average times are above 32, it is normal for the general speed to become slow.
You can reduce the average times.
8. Staircase waveform is displayed:
This phenomenon is normal because the horizontal time base is too low, and the horizontal time base may be
increased to raise the horizontal resolution, and then improve the display;
The display type may be “line”. The connection between sampling points may lead to the display of
staircase waveform. This problem can be solved by setting the display type as “point” display mode.
9. There is persist of vision in waveform display:
This phenomenon is normal because the persist time may be set too long, and the waveform persist shows
persist of vision;
The reason may be that sampling mode is set to envelope of sampling, and the sampling mode can be
modified to normal waveform for normal display