Micsig STO1152C+ STO1152C Plus digital oscilloscope

Micsig STO series user manual - Page 272

For STO1152C+.

PDF File Manual, 353 pages, Read Online | Download pdf file

STO1152C+ photo
Loading ...
Loading ...
Loading ...
Chapter 12 Serial Bus Trigger and Decode (Optional)
257
Figure 12-27 ARINC429 Trigger Mode Configuration Menu
If LABEL, SDI (source identifier), DATA or SSM (symbol/status mark) trigger are used, after selecting trigger
mode, use the pop-up virtual keyboard to modify it, enter the value, and click “Enter” on the virtual soft
keyboard to complete the setting.
Trigger configuration menu description:
a) Word start: Trigger at the word start.
b) Word stop: Trigger at the word stop.
c) LABEL: Label, triggered when the specified tag value occurs.
Loading ...
Loading ...
Loading ...