Chapter 6 Trigger System
123
6.8 Timeout Trigger
Timeout trigger happens when the time from the intersection of signal and trigger level and above (or below) the
trigger level reaches the set time, as shown in Figure 6-27:
Figure 6-27 Timeout Trigger Schematics
Timeout trigger menu descriptions are shown in the table below:
Trigger
Option
Setting
Description
Duration
Set Level