110
Logic trigger operation steps between channels:
1) Tap “Trigger” on the main menu to open the trigger menu, select logic trigger in the trigger type, and set the
logic trigger as follows, as shown in Figure 6-20:
Logic levels: CH1, CH3: High; CH2, CH4: Low; (without reference to the channel of logic operation, the
level selection is None to avoid interference to the logic operation);
Logic gate: AND;
Condition: <;
Logic time: 400ns.
Figure 6-20 Logic Trigger Setting Menu