Digital Oscilloscope Series
44
CHANNEL
OFFSET | OFST
Command/Query
DESCRIPTION
The OFFSET command allows adjustment of the
vertical offset of the specified input channel. The
maximum ranges depend on the fixed sensitivity
setting.
The OFFSET? query returns the offset value of
the specified channel.
COMMAND SYNTAX
<channel>:OFFSET <offset>
<channel>:={C1,C2,C3,C4}
<offset>:= vertical offset value with unit, see the
data sheet for details.
Note:
If there is no unit (V/mV/uV) added, it
defaults to volts (V).
If you set the offset to a value outside of the
legal range, the offset value is automatically set
to the nearest legal value. Legal values are
affected by the probe attenuation setting.
QUERY SYNTAX
<channel>:OFFSET?
RESPONSE FORMAT
<channel>:OFFSET <offset>
<offset>:= Numerical value in E-notation with
SI unit.
EXAMPLE
The following command sets the offset of
Channel 2 to -3 V.
Command message:
C2:OFST -3V
The following command sets the offset of
Channel 1 to -50 mV.
Command message:
C1:OFST -50mV
RELATED COMMANDS