Loading ...
Loading ...
Loading ...
Digital Oscilloscope Series
100
channels.
(first rising edge - last
falling edge)
LFR
Delay between two
channels.
(first falling edge - last
rising edge)
LFF
Delay between two
channels.
(first falling edge - last
falling edge)
Skew
Delay between two
channels.
(edge – edge of the same
type)
QUERY SYNTAX
<sourceA-sourceB>:MEASURE_DELY?
<type>
RESPONSE FORMAT
<sourceA-sourceB>:MEAD <type>,<value>
Model
Format of <value>
SDS1000X-
E
Numerical value
in E-notation with
SI unit, such as
1.24E-04S. Except
for PHA, it returns
as ―44.65degree‖.
others
Numerical value
in E-notation with
SI unit, such as
2.06E+01degree.
EXAMPLE
The following steps show how to get the
measured value of phase between C2 and C4.
Step 1: Send the message to set the
measurement to Phase between C2 and C4, and
then there displays a phase measurement on the
screen.
Command message:
MEAD PHA,C2-C4