ACQUIRE
ARM_ACQUISITION | ARM
Command
DESCRIPTION
The ARM_ACQUISITION command starts a
new signal acquisition.
COMMAND SYNTAX
ARM_ACQUISITION
EXAMPLE
The following steps show the effect of ARM.
Note:
INR bit 13 (8192) = Trigger is ready.
INR bit 0 (1) = New Signal Acquired.
Step 1: Set the trigger mode to single, and input
a signal which can be triggered. Once triggered,
you can see the state of acquisition changes to
stop. Send the query.
Query message:
INR?
Response message:
INR 8193(trigger ready)
Step 2: Send the query again to clear the
register.
Query message:
INR?
Response message:
INR 0
Step 3; Now, send the command to start a new
signal acquisition.
Command message:
ARM
Step 4: Send the query to see the effect of ARM.
Query message:
INR?
Loading ...