26
Maximum input voltage of the analog input
Category I 300Vrms, 400Vpk.
2.12 Use Auto
Once the oscilloscope is properly connected and a valid signal is input, tap the Auto Set button to quickly
configure the oscilloscope to be the best display effects for the input signal. While the oscilloscope in auto state, the
Auto Set button will light up .
Auto is divided into Auto Set and Auto Range. It is defaulted as Auto Set.
Auto Set — Single-time auto, and each time press “Auto”, the screen displays “Autoin the upper left corner. The
oscilloscope can automatically adjust the vertical scale, horizontal scale and trigger setting according to the
amplitude and frequency of signals, adjust the waveform to the appropriate size and display the input signal. After
adjustments, exit from the auto set, the Autoin the upper left corner disappears.
Channels may be automatically opened. Any channel greater or less than the threshold level can be opened or
closed automatically according to the set threshold level. The threshold level can be settable.