ATOx004 manual - Page 333

For ATO1004.

Loading ...
Loading ...
Loading ...
318
b) Stop bit — trigger at the stop bit of the measured signal, no matter the measured signal uses 1, 1.5, or 2
stop bits or not, the trigger will occur at the first stop bit.
c) [data] — Trigger at the specified data bit, when measured signal data bits are effective as 5 to 8bits, select
[data], and select the trigger relationship as “=”, “>” , “ <”, “≠”, then select “Trigger Data”, press data on
the touch screen, and use the pop-up virtual keyboard to modify;
d) [0:data] — the measured signal data bits is 9bits (the 9th bit is parity bit). Only when the 9th bit is 0, then
trigger. The trigger relationship, trigger data configuration are the same as those of [data] trigger;
e) [1:data] — the measured signal data bits is 9bits (the 9th bit is parity bit). Only when the 9th bit is 1, then
trigger. The trigger relationship, trigger data configuration are the same as those of [data] trigger;
f) [x:data] — the measured signal data bits is 9bits (the 9th bit is parity bit). No matter what the value of the
9th bit is, trigger at the designated data bit. The trigger relationship, trigger data configuration are the same
as those of [data] trigger;
g) Parity error — valid when there is parity check at parity bit, trigger while parity error.