resistance.
19
Series slave input during master-slave series operation.
(40 V/80 V/160 V models only)
02
Control signal common. Connected to the sense terminal
when remote sense is used.
Connected to the negative output terminal when remote
sense is not used.
04
External voltage control of the voltage output.
A voltage of 0~10 V is used to control the full-scale voltage
output (0%~100%) of the instrument.
External resistance control of the voltage output.
A resistance of 0 ~ 10 is used to control the full-scale
voltage output (0%~100%) of the instrument.
04,02
When connecting the external voltage source to the MIL
connectors, use shielded or twisted paired wiring.
06
External voltage control of the current output.
A voltage of 0~10V is used to control the full-scale current
output (0%~100%) of the instrument
External resistance control of the current output.
A resistance of 0kΩ ~ 10is used to control the full-scale
current output (0%~100%) of the instrument.
08
Current Monitor Output. Outputs the full-scale current
(0~100%) as a voltage (0V~10V).
10
Voltage Monitor Output. Outputs the full-scale voltage
(0~100%) as a voltage (0V~10V).
12
Common for status signals 14, 16, and18.
14
Turns on when CC mode is active. (photo coupled open
collector output)