SPS5000X Series User Manual
WWW.SIGLENT.COM 64
⑤ Output delay time
⑥ Slope level
When the power supply is set to CC mode priority, here is the current slope. When set
to CV priority, this is the voltage slope.
Setting method: In the List setting interface, move the cursor to here, turn the knob to
change the level, and press the knob to confirm the setting.
⑦ Number of List groups in the current setting interface, 01 means the first group
Setting method: In the List setting interface, press the key when the cursor is hidden to
switch the current group number interface.
⑧ Total number of steps, the total number of steps output by List, 05 means there are 5
sets of data
Setting method: In the List setting interface, move the cursor to here, turn the knob to
change the value, and press the knob to confirm.
V(V)
T
0
delay1 Run time1 delay2 Run time2 delay3
Run time3
delay4 Run time4 OFFON
List mode
2) Multi-channel model
To set the List output, first set the channel mode to List mode. Setting method: Press Menu
→ Configure → 10.Channel Mode to enter the setting interface. When the cursor is
hidden, press the left and right direction keys to switch the channel to be set
CH1/CH2/ CH3. CH1 means that the CH1 channel is currently selected. Press the knob and
then turn the knob to select . Press the knob. When it is displayed as List , the channel
List mode is set successfully.
For other settings, refer to the single-channel output model method. During List output, press
the knob to pause the time, and press the knob again to continue the output countdown.