5.Advanced User Guidebook
47
The description of Pass/Fail Menu is shown as the follows:
Pass/Fail test:
Detect whether the input signal is within the limits of the rule, if it exceeds limits of
the rule, it is "Fail"; otherwise it is "Pass". Also it can output fail or pass signal by
built-in and configurable output port. To run the test, read the following steps:
1. Press the Utility button, select Pass/Fail in the left menu.
2. Enable switch on: select Operator in the bottom menu, set Enable as ON in the
right menu.
3. Create rule: select Rule in the bottom menu. Select Source in the right menu;
select the desired source in the left menu. Select Horizontal or Vertical in the
right menu; turn the G knob to set Horizontal or Vertical tolerance. Select Create
in the right menu.
4. Set output type: Select Output in the bottom menu. Choose any one or two of the
options "Pass", "Fail" or "Ring". "Pass" and "Fail" are mutually exclusive
options, which could not be chosen simultaneously.
5. Begin to test: Select Operator in the bottom menu. Select Operator as Start in
the right menu, the test will begin.
6. Save rule: Select Save rule in the bottom menu. Select Save in the right menu to
enter file explorer. The path and file name are shown in the Save To: line in the
bottom. Select the save location, see "To select the path or file in file explorer" on
P43. If you want to customize the file name, select Edit Filename in the right
menu, see "To edit file names" on P43. Select OK Save in the right menu.
7. Recall rule: Select Save rule in the bottom menu. Select Recall in the right menu
to enter file explorer. Select rule file(*.PF), see "To select the path or file in file
explorer" on P43. Select the channel that you want to load to in the right menu.
Function Menu
Setting
Description
Operator
Enable
Operator
Control enable switch
Control operate switch
Output
Pass
Fail
Ring
Info
Signal tested corresponds with the rule
Signal tested not correspond with the rule
Ring when it satisfies the rule
Control the display status of info frame on the top left
Rule
Source
Horizontal
Vertical
Create
Select source
,,,
or Math
Change the Horizontal tolerance value by turning the
G knob
Change the Vertical tolerance value by turning the
G knob
Use the rule set as testing rule
Save rule
Save
Recall
Save the rule to internal/external storage
Recall a rule from internal/external storage
as the
testing rule
Loading ...
Loading ...