5.Advanced User Guidebook
34
The description of the Pulse Width Trigger Menu is as follows:
MENU
SETTING
INSTRUCTION
Trigger
Mode
Pulse Set vertical channel trigger type as pulse trigger.
Source
Channel 1 as trigger signal.
Channel 2 as trigger signal.
Channel 3 as trigger signal.
Channel 4 as trigger signal.
Coupling
AC
DC
HF
LF
Not allow DC portion to pass.
Allow all portion pass.
Not allow high frequency of signal pass and only low
frequency portion pass.
Not allow low frequency of signal pass and only high
frequency portion pass
When
Polarity
Choose the polarity
Select pulse width condition and adjust the G knob to
set time.
Mode
Holdoff
Auto
Normal
Single
Holdoff
Reset
Acquire waveform even no trigger occurred
Acquire waveform when trigger occurred
When trigger occurs, acquire one waveform then stop
100ns~10s, adjust G knob to set time interval before
another trigger occur.
Set Holdoff time as 100ns
4. Slope Trigger
Slope trigger sets the oscilloscope as the positive/negative slope trigger within the
specified time.
The description of the Slope Trigger Menu is as follows:
MENU
SETTING
INSTRUCTION
Trigger
Mode
Slope Set vertical channel trigger type as slope trigger.
Source
Channel 1 as trigger signal.
Channel 2 as trigger signal.
Channel 3 as trigger signal.
Channel 4 as trigger signal.
When
slope
Slope selecting
Set slope condition; turn the G knob to set slope time.