MAINTENANCE
327
Vertical Accuracy Calibration
Panel Operation
1. Press the Utility key.
Utility
2. Press System from the bottom
menu.
3. Press more 1 of 3 from the side
menu.
4. Press Self Cal on the side menu.
5. Press Vertical on the side menu.
6. A message appears to “Now performing
vertical calibration
CH1
Connect the CAL output to channel, then press
the Vertical key”.
7. Connect the calibration signal from the rear
panel to the Channel 1 input with a BNC cable.
CH1