Chapter 2 To Set the Vertical System RIGOL
MSO5000-E User Guide 2-3
To Adjust the Vertical Scale
Vertical scale indicates the voltage value per grid in the vertical axis of the screen. I t
is often expressed in V/ div. While you adjust the vertical scale, the display amplitude
of the waveform would enlarge or reduce. The scale information of the channel
status label (e.g. as shown in the following f igure) at the lower section of the screen
would change accordingly.
The adjustable range of the vertical scale is related to the currently set probe ratio
and input impedance. By default, the probe ratio is 1X and the input impedance is 1
MΩ. The adjustable range of the vertical scale is from 500 μV/div t o 1 0 V/ d iv.
When CH1 is turned on, you can adjust the vertical scale with the following three
methods:
Rotate Vertical
SCALE that corresponds to CH1 to adjust the vertical scale
(clockwise to reduce the scale and counterclockwise to increase).
Enable the touch screen function and adjust the vertical scale with the pinch &
stretch gesture on the touch screen. For details, refer to the "Pinch & Stretch"
section.
Enable the touch screen function, and tap the channel status label at the bottom
of the screen. The following window is displayed. Tap the
icon at the
right side of the Scale input f ield to decrease or increase the scale. You can also
tap the Scale input f ield to input a specif ic value with the displayed numeric
keypad.
In the vertical control area (Ver tical) on the front panel, press Vertical
SCALE
to adjust the vertical scale in "Coarse" (by default)
or "Fine" mode. You can also
press 1 More Fine to enable or disable f ine adjustment.
Fine adjustment: Rotate Vertical SCALE to further adjust the vertical scale
within a relatively smaller range to improve vertical resolution. I f the amplitude
of the input waveform is a little bit greater than the full scale under the current
scale and the amplitude would be a little bit lower if the next scale is used, f ine
adjustment can be used to improve the amplitude of waveform display to view
signal details.
Coarse adjustment (take counterclockwise as an example): Rotate Vertical
SCALE to set the vertical scale at 1-2-5 step, i.e. 500 μV/div, 1 mV/div, 2 mV/div,
5 mV/ div, 10 mV/div10 V/div.
Vertical Scale
Value I nput Field Decrease I ncrease