2271A
Operators Manual
70
CPS Use
To use the CPS, connect the UUT to the top test port on the CPS using the same
method as described in the Connect a UUT section.
WCaution
To prevent Product damage or damage to a gauge:
• DO NOT use PTFE tape on these connections. This prevents
a correct seal. The gauge adapter sealing system can be
hand-tight sealed up to 20 MPa (3000 psi). Wrenches or
similar tools are not required. Over tightening can cause
damage to threads or sealing faces.
• Before connection, ensure that there is an O-ring fitted to
the test port.
• Check that the sealing face of the device to be fitted is clean
and undamaged, as scratches or dents can form leak-paths.
Note
The threads on the test port, and the lower part of the gauge
adapters are left-handed.
1. Screw on the appropriate gauge adapter to the UUT, see Figure 18.
hwr034.eps
Figure 18. Screw on Gauge Adapter
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com