Automated Pressure Calibrator
Contamination Prevention System (CPS)
67
Contamination Prevention System (CPS)
Note
Maximum working pressure (MWP) of the CPS is 20 MPa
(3,000 psi.)
The Contamination Prevention System (CPS) is a Product accessory that
protects the Product from contamination from the UUT.
The CPS accomplishes this through these methods:
The CPS is electrically connected to the Product. On downward pressure
excursions, the Product releases the pressure through its vent valve. While
the precise pressure control is done in the Product, the major change in
pressure is accomplished through the Product. This results in uni-directional
flow between the Product and the CPS.
Substances from the UUT drop into a sump system that uses gravity to trap
any liquids. The sump is routinely purged when the system is vented.
The CPS includes both a 25 micron screen and coalescing filter.
WCaution
To prevent damage to the Product:
Do not use the CPS if it is not turned on in the Settings.
Service the filters regularly. Service frequently if the UUT
has a large amount of fluid in it or is very dirty.
Do not let the fluid in the bottle fill to the top.
WWarning
To prevent injury:
Do not use CPS with flammable or combustible liquids.
Only use the CPS with a pressure controller rated for no
greater than the MWP of the CPS 20 MPa (3,000 psi).