2271A
Operators Manual
34
Connect a UUT
Connect the UUT to the top test port on the Product using the method described
below:
W Caution
To prevent Product damage or damage to a gauge:
DO NOT use PTFE tape on these connections. This prevents
a correct seal. The Gauge Adapter sealing system can be
hand-tight sealed up to 20 MPa (3000 psi). Wrenches or
similar tools are not required. Over tightening can cause
damage to threads or sealing faces.
Before connection, ensure that there is an O-ring fitted to
the test port.
Check that the sealing face of the device to be fitted is clean
and undamaged, as scratches or dents can form leak-paths.
Note
The thread on the test port, and the lower part of the gauge adapters
is LEFT-HANDED. The following procedure details the correct
method for mounting devices using these adapters:
1. Screw the appropriate gauge adapter fully on to the UUT, see Figure 6.
hwr029
Figure 6. Screw on Gauge Adapter
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com