4.Advanced User Guidebook
69
Remove Remove the chosen measure type
Remove All
Remove all the measures
Show
CH1
ON
OFF
Show all the measures of CH1 on the screen
Hide the window of CH1 measures
Show
CH2
ON
OFF
Show all the measures of CH2 on the screen
Hide the window of CH2 measures
Measure
Only if the waveform channel is in the ON state, the measurement can be performed.
The automatic measurement can not be performed in the following situation: 1) On
the saved waveform. 2) On the Dual Wfm Math waveform. 3) On the Video trigger
mode.
On the Scan format, period and frequency can not be measured.
Measure the period, the frequency of the CH1, following the steps below:
1. Push the Measure button to show the automatic measurement function menu.
2. Select Add in the bottom menu.
3. In the right menu, select CH1 in the Source menu item.
4. In the left Type menu, turn the M knob to select Period.
5. In the right menu, select Add. The period type is added.
6. In the left Type menu, turn the M knob to select Frequency.
7. In the right menu, select Add. The frequency type is added.
The measured value will be displayed at the bottom left of the screen automatically
(see Figure 4-13).
Figure 4-13 Automatic measurement