4.Advanced User Guidebook
50
the horizontal scale and position.
The Vertical Scale and the Vertical Position knobs of Channel 2 are used to set
the vertical scale and position continuously.
The following functions can not work in the XY Format:
Reference or digital wave form
Cursor
Trigger control
FFT
Operation steps:
1. Push the Display button.
2. Select XY Mode in the bottom menu. Select Enable as ON in the right menu.
3. To make the XY view full screen, select Full Screen as ON in the right menu.
Counter
It is a 6-digit single-channel counter. The counter can only measure the frequency of
the triggering channel. The frequency range is from 2Hz to the full bandwidth. Only if
the measured channel is in Edge mode of Single trigger type, the counter can be
enabled. The counter is displayed at the bottom of the screen.
Operation steps:
1. Push Trigger Menu button, set the trigger type to Single, set the trigger mode to
Edge, select the signal source.
2. Push the Display button.
3. Select Counter as ON or OFF in the bottom menu.
How to Save and Recall a Waveform
Push the Save button, you can save the waveforms, configures, screen images, record
or clone the waveform.
The description of the Save Function Menu is shown as the following table:
Function Menu
Setting
Description
Type
Wave
Configure
Image
Record
Clone
Choose the saving type.
About the Record type, see "How to
Record/Playback Waveforms" on P57.
About the Clone type, see How to Clone
a waveform” on P61.
When the type is Wave, the menu shows as following: