4.Advanced User Guidebook
51
Type
Wave
Format
(Right
menu)
For internal storage, only BIN can be
selected. For external storage, the format
can be BIN, TXT or CSV.
Source
CH1
CH2
Math
All
Choose the waveform to be saved. (If
certain channel is off, the corresponding
menu item will be disabled.)
(Choose All to save all the waveforms that
are turned on. You can save in
to the
current internal object address, or into USB
storage as a single file.)
Object & Show
Object 0 - 49
Choose the address which the waveform is
saved to or recall from.
Show
ON
OFF
Recall or close the waveform stored in the
current object address.
When the show is
ON, if the current object address has been
used, the stored waveform will be shown,
the address number and relevant
information will be displayed at the top left
of the screen; if the address is empty, it
will prompt "None is saved".
Close All
Close all the waveforms stored in the
object address.
Save
Save the waveform of the source to the
selected address. Whatever the Type of
save menu is set, you can save the
waveform by just pressing the Copy panel
button in any user interface.
Select Type in the bottom menu,
in the
right Format
menu, you can select the
storage format.
Storage
Internal
External
Save to internal storage or USB storage.
When External is selected, save the
waveform according to the current record
length (see "Record Length menu" on P46);
the file name is editable. The BIN
waveform file could be open
by OWON
waveform analysis software (on the
supplied CD).
When the type is Configure, the menu shows as following:
Configure
Setting0
…..
Setting19
The setting address
Save
Save the current oscilloscope configure to
the internal storage
Loading ...
Loading ...