4.Advanced User Guidebook
18
How to Set the Vertical System
The VERTICAL CONTROLS includes three menu buttons such as CH1, CH2 and
Math, and four knobs such as Vertical Position, Vertical Scale for each channel.
Setting of CH1 and CH2
Each channel has an independent vertical menu and each item is set respectively based on
the channel.
To turn waveforms on or off (channel, math)
Pushing the CH1, CH2, or Math buttons have the following effect:
• If the waveform is off, the waveform is turned on and its menu is displayed.
• If the waveform is on and its menu is not displayed, its menu will be displayed.
• If the waveform is on and its menu is displayed, the waveform is turned off and its menu
goes away.
The description of the Channel Menu is shown as the following list:
Function
Menu
Setting
Description
Coupling
DC
AC
GROUND
Pass both AC and DC components of the input signal.
Block the DC component of the input signal.
Disconnect the input signal.
Inverted
ON
OFF
Display inverted waveform.
Display original waveform.
Probe
Attenu
0.001X
to
1000X
Step by 1 2 – 5. Match this to the probe attenuation
factor to have an accurate reading of vertical scale.
MeasCurr
YES
NO
If you are measuring current by probing the voltage
drop across a resistor, choose YES.
A/V (mA/V)
V/A (mV/A)
Turn the M knob to set the Amps/Volts ratio. The
range is 100 mA/V - 1 KA/V.
Amps/Volts ratio = 1/Resistor value
Volts/Amp ratio is automatically calculated.
Limit
Full band
20M
Get full bandwidth.
Limit the channel bandwidth to 20MHz to reduce
display noise.
1. To set channel coupling
Taking the Channel 1 for example, the measured signal is a square wave signal containing
the direct current bias. The operation steps are shown as below:
(1) Push the CH1 button to show the CH1 SETUP menu.
(2) Select Coupling in the bottom menu.