SIGLENT
SDS1000CML+/SDS1000DL+ User Manual 59
8. Set up Menu Display
Press the “Menu Display” option button to select “2 sec”, “5sec”, “10sec”,
“20sec” or “Infinite” to set menu display time on screen.
9. Set Skin
Press the “skin” option button or turn the “Universal” knob to select “Classical”,
“Modern”, “Traditional” or “Succinct”.
2.10.1 X-Y Format
Use the XY format to analyze phase differences, such as those represented by
Lissajous patterns. The format plots the voltage on channel 1 against the voltage
on channel 2, where channel 1 is the horizontal axis and channel 2 is the vertical
axis. The oscilloscope uses the untriggered Sample acquisition mode and
displays data as dots.
XY mode has a breakthrough that traditional oscilloscopes restrict sampling rate
at 1MSa/s,
support 25KSa/s~250MSa/s (1-2.5-5 sequence) adjusted ;
Operation steps :
●Channel 1 “Volt/div” and vertical “POSITION” set up the horizontal scale and
position.
●Channel 2 “Volt/div” and vertical “POSITION” set up the horizontal scale and
position.
●Turn the “S/div” knob to adjust the sampling rate
●The following functions are forbidden at XY display form:
●Benchmark wave form and wave mathematic
●Cursor
●Auto(resets display format to YT)
●Trigger Control
●Horizontal Position Knob
●Vector Display Type
●Scan Display