Loading ...
Loading ...
Loading ...
SIGLENT
SDS1000CML+/SDS1000DL+ User Manual 59
8. Set up Menu Display
Press the Menu Display option button to select 2 sec, 5sec, 10sec,
20sec or Infinite to set menu display time on screen.
9. Set Skin
Press the skin option button or turn the Universal knob to select Classical,
Modern, Traditional or Succinct.
2.10.1 X-Y Format
Use the XY format to analyze phase differences, such as those represented by
Lissajous patterns. The format plots the voltage on channel 1 against the voltage
on channel 2, where channel 1 is the horizontal axis and channel 2 is the vertical
axis. The oscilloscope uses the untriggered Sample acquisition mode and
displays data as dots.
XY mode has a breakthrough that traditional oscilloscopes restrict sampling rate
at 1MSa/s
support 25KSa/s250MSa/s (1-2.5-5 sequence) adjusted ;
Operation steps
Channel 1 “Volt/div” and vertical “POSITION” set up the horizontal scale and
position.
Channel 2 “Volt/div” and vertical “POSITION” set up the horizontal scale and
position.
Turn the “S/div” knob to adjust the sampling rate
The following functions are forbidden at XY display form:
Benchmark wave form and wave mathematic
Cursor
Autoresets display format to YT
Trigger Control
Horizontal Position Knob
Vector Display Type
Scan Display
Loading ...
Loading ...
Loading ...