50 SDS1000CML+/SDS1000DL+ User Manual
2.9 Acquiring Signals System
Showing as the follow picturethe “ACQUIRE button for Acquiring Signals system
is at the menu.
Picture 2.9-1
Table 2-28 The Function manual of Acquiring Signals:
Option
Setting
Introduction
Acquisition
Sampling
Use for sampling and accurately display most
of the waveform.
Peak
Detect
Detect the noise and decrease the
possibility of aliasing.
Average
Use to reduce random or uncorrelated noise
in the signal display.
Averages
4, 16,32
,64,128,
256
Select number of averages.
Sinx/x
On
off
Use sin interpolation
Use linear interpolation
Mode
Equ time
Real time
Set the Sampling mode to Equ time.
Set the Sampling mode to Real time.
Sa Rate
Display system sampling rate.
When you acquire a signal, the oscilloscope converts it into a digital form and
displays a waveform. The acquisition mode defines how the signal is digitized and
the time base setting affects the time span and level of detail in the acquisition.
Sampling: In this acquisition mode, the oscilloscope samples the signal in
evenly spaced intervals to construct the waveform. This mode accurately
represents signals most of the time.
Advantage: You can use this mode to reduce random noise.
Disadvantage: This mode does not acquire rapid variations in the signal that may