Binder VD 115 Vacuum Drying

User Manual - Page 162

For VD 115.

PDF File Manual, 179 pages, Read Online | Download pdf file

VD 115 photo
Loading ...
Loading ...
Loading ...
VD (E3.1) 10/2020 page 162/179
Chamber size
23
56
115
Different electrical data for VD-UL constructed for the USA and Canada
(model versions VD023UL-120V, VD056UL-120V, VD115UL-120V)
Nominal voltage
(+/-10%)
at 50 Hz power frequency
V
100-120
100-120
100-120
at 60 Hz power frequency
V
100-120
100-120
100-120
Nominal current
A
7,5
11,7
13,4
Chamber fuse 5 x 20 mm / 250V / time-lag T
A
10
2 x 16
--
Chamber fuse 6.3 x 32 mm / 250V / time-lag
T
A -- -- 2 x 20
Power plug
NEMA
5-15P
5-20P
5-20P
Power plug of pump module (option)
NEMA
5-15P
5-15P
5-15P
Environment-specific data
Noise level VD (mean value)
dB (A)
40
40
40
Energy consumption
at 100 °C / 212 °F
Wh/h
140
180
230
at 220 °C / 428 °F
Wh/h
370
560
750
All technical data is specified for unloaded chambers with standard equipment (with aluminum racks) at
an ambient temperature of +22 °C +/- 3 °C / 71.6 °F +/- 5.4 °F and a power supply voltage fluctuation of
+/- 10 %. Technical data is determined in accordance to BINDER Factory Standard Part 2:2015 and DIN
12880:2007.
All indications are average values, typical for chambers produced in series. We reserve the right
to change technical specifications at any time.
27.4 Equipment and options (extract)
To operate the chamber, use only original BINDER accessories or accessories / components
from third-party suppliers authorized by BINDER. The user is responsible for any risk arising
from using unauthorized accessories.
Regular equipment
RD4 microprocessor program controller with LED display
Ethernet interface for computer communication
USB interface
Safety controller (safety device class 2 according to DIN 12880:2007)
Inert gas connection / ambient air inlet GAS/AIR”, adapter with hose olive 8 mm, with fine-dosing valve
for aeration / inert gas
Measuring connection (DN 16), rear
Safety glass panel
Options / accessories
Expansion racks, aluminum or stainless steel 1.4571
Microprocessor program controller MB2 with touchscreen
Lockable door
Interior LED light
Analog outputs 4-20mA for temperature and pressure
Additional inert gas connection / ambient air inlet “GAS/AIR 2, adapter with hose olive 8 mm, with fine-
dosing valve for aeration / inert gas
Additional universal access port 40 mm / 1.57 in
FKM door gasket (temperature resistant up to 150 °C / 302 °F)
Object temperature display with flexible Pt 100 temperature sensor