Binder VD 115 Vacuum Drying

User Manual - Page 113

For VD 115.

PDF File Manual, 179 pages, Read Online | Download pdf file

VD 115 photo
Loading ...
Loading ...
Loading ...
VD (E3.1) 10/2020 page 113/179
16.7.2 Set-point ramp and set-point step
You can define the type of temperature and pressure transitions for each individual program section.
“Ramp” mode: Gradual changes of temperature and pressure
The set-point of a given program section functions as the section’s start temperature. During the section’s
duration, the set-point gradually passes to the set-point of the subsequent program section. The actual
value follows the continually changing set-point.
If the last program section is in “ramp” mode and the setpoint shall change within this section, then you
must program an additional section (with the shortest possible section duration) to provide the target
temperature of the last program section. Otherwise, the setpoint would remain constant during the
section’s duration.
Programming in the “ramp” mode allows all kinds of temperature and humidity transitions:
Gradual changes of temperature and pressure
The setpoint changes its value gradually during the entered section duration. The actual value follows
the continually moving set-point at any time.
Program sections with constant temperature and pressure
The setpoints (initial values) of two subsequent program sections are identical; so the temperature and
pressure remain constant during the entire duration of the first program section.
Sudden changes of temperature and pressure
Steps can be programmed in ramp mode as temperature or humidity changes (ramps) that occur
during a very short interval. If the duration of this transitional program section is very short (minimum
entry 1 sec), the temperature or humidity change will proceed rapidly within the minimum amount of
time.
“Step” mode: Sudden changes of temperature and pressure
The set-point of any program section functions as the section’s target value. At the start of the program
section, the unit heats up and evacuates / ventilates the chamber with the maximum speed to reach the
entered value; and then it holds it for the remaining section time. Therefore the set-point temperature
remains constant for the section’s duration. These changes occur rapidly within the minimum amount of
time (minimum entry: 1 second).
Programming in the “step” mode allows only two kinds of temperature and pressure transitions:
Programming gradual changes of temperature and humidity (ramps) is impossible in the “step” mode
Program sections with constant temperature and pressure
The setpoints (target values) of two subsequent program sections are identical; so the temperature
and pressure remain constant during the entire duration of the first program section.
Sudden changes of temperature and pressure
The entered setpoint of the section is reached as fast as possible and then held constant for the
remaining section duration.
Selecting the setting “Ramp” or “Step”
Section view (partial view).
In the field “Course” select the desired setting
“Ramp” or “Step”.
Loading ...
Loading ...
Loading ...