Binder VD 115 Vacuum Drying

User Manual - Page 160

For VD 115.

PDF File Manual, 179 pages, Read Online | Download pdf file

VD 115 photo
Loading ...
Loading ...
Loading ...
VD (E3.1) 10/2020 page 160/179
27. Technical description
27.1 Factory calibration and adjustment
This chamber was calibrated and adjusted in the factory. Calibration and adjustment were performed
using standardized test instructions, according to the QM DIN EN ISO 9001 system applied by BINDER
(certified since December 1996 by TÜV CERT). All test equipment used is subject to the administration of
measurement and test equipment that is also a constituent part of the BINDER QM DIN EN ISO 9001
systems. They are controlled and calibrated to a DKD-Standard at regular intervals.
Factory adjustment was done in the center of the usable volume and under vacuum conditions. the
sensor is fixed in the middle of the expansion rack in a way ensuring good thermal conductivity (heat
conduction). Measuring is performed in equilibrated state.
NOTICE
Danger of invalid calibration caused by modified thermal conductivity when
changing between aluminum and stainless-steel racks.
Undefined heating performance.
Do NOT change between aluminum and stainless-steel racks.
Operate the chamber with the supplied expansion racks only.
27.2 Over current protection
The chambers are protected by one or two miniature fuses against over current, accessible from the
outside. The miniature fuses are located at the rear of the chamber next to the power cable connection.
Each fuse holder is equipped with a fuse clip 5mm x 20 mm (cUL-Version 6,3x32 mm). A fuse may be
replaced only with a substitute of the same ratings. Refer to the technical data of the respective device
type.
27.3 VD / VD-UL technical data
Chamber size
23
56
115
External dimensions
Vacuum oven
Width, net
mm / inch
523 / 20.59
638 / 25.12
743 / 29.25
Height, gross (including feet)
mm / inch
720 / 28.35
837 / 25.08
964 / 37.95
Depth, net
mm / inch
490 / 19.29
540 / 21.26
660 / 25.98
Depth, gross (including door handle,
connections)
mm / inch 596 / 23.46 645 / 25.39 764 / 30.08
Window width
mm / inch
265 / 10.43
380 / 14.96
486 / 19.13
Window height
mm / inch
265 / 10.43
380 / 14.96
486 / 19.13
Pump module (option)
Width, net
mm / inch
523 / 20.59
638 / 25.12
743 / 29.25
Height, net
mm / inch
705 / 27.76
705 / 27.76
705 / 27.76
Depth, net
mm / inch
491 / 19.33
539 / 21.22
659 / 25.94
Depth, gross (including door handle,
connections)
mm / inch 546.5 / 21.52 594.5 / 23.41 714.5 / 28.13
Chamber with optional pump module
Width, net
mm / inch
523 / 20.59
638 / 25.12
743 / 29.25
Height, net
mm / inch
1425 / 56.10
1542 / 60.71
1707 / 67.20
Depth, net
mm / inch
720 / 28.35
837 / 25.08
964 / 37.95
Depth, gross (including door handle,
connections)
mm / inch 596 / 23.46 645 / 25.39 764 / 30.08
Loading ...
Loading ...
Loading ...