4. SPECIFICATIONS
4.1 Technical Information
Channel Four, isolated between channels
Power Supply Four 1.5V AA Batteries
Waveform Asymmetrical Biphasic Square Pulse
Pulse Amplitude Adjustable, 0 – 80 mA at 500-ohm load, each channel
Voltage 0 to 40V (Load: 500 ohm)
Pulse Rate Adjustable, 2 to 150 Hz (0.5 to 150 Hz for Burst Only)
Pulse Width Adjustable from 50 to 300 microseconds, 10 us/step
Contraction (ON) Time Adjustable, 2 – 60 seconds, continuous
Relaxation (OFF) Time Adjustable, 0 – 60 seconds, continuous
Ramp Time (up and
down)
Adjustable, 1 – 8 seconds
Size 5.5 in (L) x 3.1 in (W) x 1.1 in (H)
Weight 1 lb.
Operating Condition Temperature: 32° F – 104° F (0°C – 40°C)
Relative Humidity: 30% – 75%
Atmosphere Pressure: 700nPa – 1060hPa
Low Battery Indicator A low battery indicator will show up when the battery is low
Timer Two Adjustable Timers, from 1 – 60 minutes or Continuous.
Adjustable in 1 minute steps from 1 – 15 minutes and then
5 minute steps from 15 – 60 minutes. To increase or decrease
timer quicker, press and hold the increase or decrease button.
Patient Compliance
Meter
The device can store 60 sets of usage records with a total
treatment time of 999 hours.
Standard Deviation
Remark
There may be up to a +/- 5% tolerance of all parameters and
a +/- 20% tolerance of amplitude & voltage.
Service Life of the Device 3 years
Service Life of the
Electrodes
Electrodes can be cleaned and reused for up to 1~15 times
Service Life of the
Battery
New batteries will for approx. 20 times (when used for 30
minutes a day, N mode, in half of the maximum intensity).
Applied Part Electrode