TROUBLESHOOTING
LCD Display Possible Cause Solution
Low battery Please check if the battery has
been loaded correctly. Replace
the battery.
Product doesn’t
turn off
Internal malfunction Remove and replace battery
Patient or thermometer was
moved from a cold environment.
Thermometer probe was recently
wet cleaned.
Wait at least 20 minutes before
retaking temperature.
Wait at least 5 minutes after
cleaning before retaking
temperature.
Patient or thermometer was
moved from a warm environment.
Patient was laying on the same
side where measurement was
taken.
Wait at least 20 minutes
before retaking temperature.
Measure temperature from
the opposite side.
Internal malfunc tion Restart t hermometer.
The ERR message always comes
with a number. If restarting the
thermometer does not solve the
issue, note this number, call your
authorized Kaz Service Center to
inform them about the ERR.
PRODUCT SPECIFICATIONS
Displayed temperature range: 34-42.2 °C (93.2-108 °F)
Measurement time: 1 second
Operating ambient temperature range: 10-40 °C (50-104 °F)
Display resolution: 0.1 °C or °F
Calibration Accuracy: ± 0.1 °C (0.2 °F)
Battery life: 3 years or 500 measurements
Type BF equipment
Subject to change without notice.
This product conforms to the provisions of the EC directive 93/42/EEC
(Medical Device Directive).
MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC.
For detailed description of EMC requirements please contact an authorized local
Service Centre (See Authorized Service Centers card).
Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL
EQUIPMENT.
Thermometer was moved during
temperature taking process
Thermometer was not in the
correct position during the
temperature taking process
Short measurement time
Start button was pressed too
slow
Hold thermometer still and
retake measurement
Make sure yellow ring of
thermometer is behind ear lobe
and retake measurement.
Make sure that the Start Button
is kept pressed until the beep
sounds.
Retake the temperature and
ensure the start button is pressed
down steadily