Specifications
Pressure Specification
1-year Specification ................................................................ 0.02 % of full scale
Control Specification ............................................................... 0.005 % full scale minimum
Temperature Compensation ................................................... 15 °C to 35 °C (59 °F to 95 °F) to rated accuracy
Note: For temperatures from -10 °C to +15 °C and 35 °C to
50 °C, add 0.04 % of full scale
Electrical Specification
All specifications are valid to 110 % of range, except 24 mA source and simulate which are valid to 100 % of range.
Ranges
mA ..................................................................................... 0 mA to 24 mA
Volts .................................................................................. 0 V dc to 30 V dc
Resolution
mA Ranges ....................................................................... 1 μA
Voltage Range ................................................................... 1 mV
Accuracy ................................................................................. 0.01 % ±2 LSD all ranges (at 23 °C ±5 °C)
Temperature Compensation ................................................... 20 ppm of full scale /°C from -10 °C to +18 °C and 28 °C to