Loading ...
Loading ...
Loading ...
![](https://files.manualsfile.com/33456604-729-30g-fc/bg30.png)
729/729FC
Users Manual
42
Set Fixed mA Output
Enter the necessary mA value to output from the connected
transmitter.
• Push B (Step Up 4mA) to increase mA output in
4 mA steps.
• Push C (Step Down 4mA) to decrease mA output
in 4 mA steps.
Re-range Transmitter
Change the Upper Range Value (URV) and the Lower
Range Value (LRV) of the connected transmitter from this
screen.
1. Enter the Upper Range Value (PV URV) necessary.
2. Push E to select.
3. Push C (Send) to change the URV.
4. Enter the Lower Range Value (PV LRV) necessary.
5. Push E to select.
6. Push C (Send) to change the LRV.
Device Diagnostics
Select Device Diagnostics to send the self-diagnostics
command to the connected transmitter and view the
transmitter Self Test result.
Calibrate (Ad hoc)
Calibrate (Ad Hoc) calibrates a connected HART device
without a downloaded task.
The opening screen shows the task type and
source/measure configuration for the device. The
transmitter HART configuration is copied into this
configuration screen. Correct as needed for your test.
1. Push B (PV Calibrate) to change the measure
from mA to the PV source value of the transmitter.
2. Push C (Disable/Enable Loop Power).
3. Push D (Continue) to continue to second
configuration screen. These choices are shown:
• Tolerance of Range
• Test Strategy
• Auto Settling Time
• Transfer Function
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com
Loading ...
Loading ...
Loading ...