Automatic Pressure Calibrator
The Display
13
RTD Connection
The display can show RTD temperature measurements in
the selected unit (C/F) as the third-function. When the
temperature unit field highlights, push the softkey to toggle
the temperature setting. The ohms reading shows.
Note
The Product supports PT100-385 RTD only.
Pressure Module Connection
The Product automatically detects external pressure module
installation or removal. Move the cursor to the unit type and
push E to configure the pressure-unit type (it can be
different from the pressure unit of internal pressure). Change
the unit type of the internal pressure to automatically change
the unit type of the pressure module.
Figure 6 shows a pressure module as the third function.
When you plug in an external pressure module, the
pressure-unit type defaults to the internal pressure unit
type.
When you remove or install an external pressure
module, the zero offset resets.
idj020.bmp
Figure 6. Triple-Function Display with External Pressure
Module Connected