Automatic Pressure Calibrator
Maintenance
57
Maintenance
XW Warning
To prevent possible electrical shock, fire, or
personal injury:
Have an approved technician repair the
Product.
Do not operate the Product with covers
removed or the case open. Hazardous
voltage exposure is possible.
Remove the input signals before you
clean the Product.
Use only specified replacement parts.
Run “Exhaust” before you open the battery
door.
Clean the Product
Clean the Product and pressure modules with a soft cloth
dampened with water or water and mild detergent.
WCaution
To prevent possible damage to the Product,
do not use solvents or abrasive cleansers.
Clean the Pump Valve
1. Remove the battery door (see Battery Replacement)
and locate the pump valve caps, see Figure 23 ().
2. Use a small screwdriver to remove the two valve
retention caps located in the oval shaped opening on
the underside of the Product.
3. Gently remove the spring and O-ring assembly.
4. Set aside the valve assemblies and clean out the
valve body with a cotton swab soaked in isopropyl
alcohol (IPA).
5. Repeat this process several times with a new cotton
swab until there is no sign of residue.
6. Run the pump for a few seconds.
7. Clean the O-ring assembly and O-ring on the
retention caps with IPA and inspect the O-rings
closely for any cuts, nicks, or wear. Replace if
necessary.