Automatic Pressure Calibrator
Tasks Menu
27
Pressure Transmitter (P/I)
Use this function to calibrate a P/I transmitter (pressure to
current) with current measurement. See Figure 11 for
connections:
1. From page 1 of the Tasks menu, push the arrows to
highlight Pressure Transmitter (P/I).
2. Push E or R.
3. From this screen, use the arrows, numeric keypad,
and E to change the values of the Source
Pressure and Measure Current.
4. Push B (PV Calibrate) if HART is enabled and
PV calibration is necessary.
5. Push C to turn on (or off) 24 V loop power.
6. Once these values are specified, push D
(Continue).
7. From this second screen, use the arrows, numeric
keypad, and E to change the values of the
Tolerance of Range, Test Strategy, Auto Settling
Time, and Transfer Function.
8. Push A to change the Transfer Function to
Square Root or Linear.
9. To select the Test Strategy, highlight the parameter
and push R. Highlight the applicable listed test
strategy and push E.
10. Push B (Save as Customized) to store this
task for later use.
11. Once you enter the parameters, push D
(Continue).
12. Push B (Manual Test) or D (Auto Test).
13. Push C to adjust. Go to 0%, 50%, and 100%.
14. Push A (As Left) to do the As Left test.