Automatic Pressure Calibrator
HART Menu
43
These softkeys are active:
A – Linear/Square Root Selection
B – Save as Customized
C – Leak Test
D – Continue
The next screen starts the As Found Calibration. These
softkeys are active:
A – Abort
B – Manual Test
D – Auto Test
Test results show after all strategy points have recorded the
measurements. Measurements in specification show as
black. Measurements out of specification show as red.
1. Push D (Done) or A (Abort).
2. Enter the Tag ID, serial number of the device, and the
User ID.
3. Push D (Done)
Adjust
Push B (HART Adjust). These choices are shown:
PV Zero Trim
mA Output Trim
Trim to Applied Values
PV Zero Trim
To set the digital pressure value of the Product to zero:
1. Push C (Trim Zero).
2. Push D (Done).
mA Output Trim
From the mA output trim screen, you can fetch the current
PV value and send this value to the transmitter to adjust the
PV to the correct mA output.
1. Starting with the 4 mA range, push A (Fetch) to
get the current measured mA value.
2. Push B (Trim 4mA) to send the measured value
for correction.
3. Push C (Next Point). The Product moves to the
20 mA range.
4. Repeat steps 1, 2, and 3.
5. When complete, push Cto return to the previous
point or D (Done).