Automatic Pressure Calibrator
Maintenance
67
Table 7. Error Messages (cont.)
Error Message Explanation Potential Root Cause
Supply Pressure Too Low
Failed to create supply pressure for the
current setpoint.
1. The pressure volume is too large, reduce
the volume and try again.
2. The leakage of pressure volume is too
large. Test the leakage and try again.
3. Ambient pressure is lower than the negative
gauge setpoint. Change the negative gauge
setpoint to a reasonable value at current
ambient pressure.
4. The unit needs service.
Value out of range Value input is out of range. Try to configure a valid value.
Function not available
Selected function is not available for the
current state.
Illegal Operation Illegal operation for the current state.