8845A/8846A
Users Manual
4-4
Measuring Voltage
The Meter is capable of measuring up to 1000 V dc, 750 V ac (8845A) or 1000 V ac
(8846A).
WCaution
To avoid blowing the current fuses and possible damage to
other equipment, do not apply voltage to the meter’s input until
the test leads are properly connected to the input, and the
proper voltage function selected.
Measuring DC Voltage
To make a dc voltage measurement:
1. Press E.
The dc voltage icon | will appear to the right of the displayed value as shown
below.
-0.001698
V
F1 F2 F3 F4 F5
caw021.eps
2. Connect the test leads to the Meter’s inputs as shown in Figure 4-1.
3. Connect the test leads to the circuit and read the measured voltage in the Meter’s
display.
2
INPUT SENSE
HI
LO
FRONT
REAR
10 A
1V
100
mA
4
W
HI
LO
2W/4W
V
300V
FUSED
1000 V CAT I
600V CAT II
Circuit
Under
Test
caw019.eps
Figure 4-1. Input Connections for Voltage, Resistance, and Frequency Measurements.
Function modifiers:
FILTER An 8 Hertz low-pass filter for rejecting noise and signals above 8 Hz
is inserted to stabilize measurements. The filter is active when this soft
1.888.610.7664 sales@GlobalTestSupply.com