Preparing the Meter for Operation
Power Considerations 2
2-5
Table 2-1. Line Voltage to Fuse Rating
Line Voltage Selection Fuse Rating Fluke Part No.
100 0.25 A, 250 V (slow blow) 166306
120 0.25 A, 250 V (slow blow) 166306
220 0.125 A, 250 V (slow blow) 166488
240 0.125 A, 250 V (slow blow) 166488
120
caw0201f.eps
Figure 2-1. Replacing the Line Fuse
Current-Input Fuses
The 100 mA and 10 A inputs are protected by user-replaceable fuses.
The 100 mA input is protected by a fuse (F2) rated at 440 mA, 1000 V (fast
blow), 10,000 A minimum breaking capacity (Fluke Part No. 943121).
The 10 A input is protected by a fuse (F1) rated at 11 A, 1000 V (fast blow),
10,000 A minimum breaking capacity (Fluke Part No. 803293).
W Warning
For protection against fire or arc flash, replace a blown fuse
only with one from Fluke.
To test for a blown Current Input Fuse:
1. With the Meter powered up, plug a test lead into the VGER connector.
2. Press N.
3. Insert the other end of the test lead into the 100 mA input connector.
If the fuse is good, the Meter will read less than 200 . If the fuse is blown, the
Meter will read overload.
4. Remove the probe from the 100 mA connector and insert into the 10 A
connector.