1-14
Max. Lead Resistance (4-wire ohms) .................. 10 % of range per lead for 100 , 1 k ranges. 1 k per lead on all
other ranges.
Input Protection .................................................... 1000 V on all ranges.
Input Characteristics
Resolution
Range
Full Scale
(6½ Digits)
4½ Digits 5½ Digits 6½ Digits
Source Current
10
[1]
10.00000 1 m 100 µΩ 10 µΩ
5 mA
100 100.0000 10 m 1 m 100 µΩ
1 mA
1 k 1.000000 k 100 m 10 m 1 m
1 mA
10 k 10.00000 k 1 100 m 10 m 100 µA
100 k 100.0000 k 10 1 100 m 10 µA
1 M 1.000000 M 100 10 1 10 µA
10 M 10.00000 M 1 k 100 10 1 µA
100 M 100.0000 M 10 k 1 k 100 1 µA || 10 M
1.0 G
[1]
1.000000 G 100 k 10 k 1 k 1 µA || 10 M
[1] 8846A Only
8846A Accuracy
Accuracy is given as ± (% measurement + % of range)
Range
24 Hour
(23 ± 1 °C)
90 Days
(23 ± 5 °C)
1 Year
(23 ± 5 °C)
Temperature
Coefficient/ °C
Outside 18 to 28 °C
10 0.003 + 0.01 0.008 + 0.03 0.01+ 0.03 0.0006 + 0.0005
100 0.003 + 0.003 0.008 + 0.004 0.01 + 0.004 0.0006 + 0.0005
1 k 0.002 + 0.0005 0.008 + 0.001 0.01 + 0.001 0.0006 + 0.0001
10 k 0.002 + 0.0005 0.008 + 0.001 0.01 + 0.001 0.0006 + 0.0001
100 k 0.002 + 0.0005 0.008 + 0.001 0.01 + 0.001 0.0006 + 0.0001
1 M 0.002 + 0.001 0.008 + 0.001 0.01 + 0.001 0.001 + 0.0002
10 M 0.015 + 0.001 0.02 + 0.001 0.04 + 0.001 0.003 + 0.0004
100 M 0.3 + 0.01 0.8 + 0.01 0.8 + 0.01 0.15 + 0.0002
1 G 1.0 + 0.01 1.5 + 0.01 2.0 + 0.01 0.6 + 0.0002
8845A Accuracy
Accuracy is given as ± (% measurement + % of range)
Range
24 Hour
(23 ± 1 °C)
90 Days
(23 ± 5 °C)
1 Year
(23 ± 5 °C)
Temperature
Coefficient/ °C
Outside 18 to 28 °C
100 0.003 + 0.003 0.008 + 0.004 0.01 + 0.004 0.0006 + 0.0005
1 k 0.002 + 0.0005 0.008 + 0.001 0.01 + 0.001 0.0006 + 0.0001
10 k 0.002 + 0.0005 0.008 + 0.001 0.01 + 0.001 0.0006 + 0.0001
100 k 0.002 + 0.0005 0.008 + 0.001 0.01 + 0.001 0.0006 + 0.0001
1 M 0.002 + 0.001 0.008 + 0.001 0.01 + 0.001 0.0010 + 0.0002
10 M 0.015 + 0.001 0.02 + 0.001 0.04 + 0.001 0.0030 + 0.0004
100 M 0.3 + 0.01 0.8 + 0.01 0.8 + 0.01 0.1500 + 0.0002
Additional Ohms Errors
Digits NPLC Additional Noise Error
100 0 % of range
10 0 % of range
1 0.001 % of range
5½ 0.2
0.001 % of range ± 20 m
4½ 0.02
0.01 % of range ± 20 m