FLUKE 8846A/SU 120V Digital Precision Multimeter with Software and Cable, 6.5

User Manual for the Fluke 8846A/SU 120V Digital Precision Multimeter with Software and Cable, 6.5 - Page 15

For 8846A/SU 120V.

PDF File Manual, 82 pages, Read Online | Download pdf file

8846A/SU 120V photo
Loading ...
Loading ...
Loading ...
Introduction and Specifications
General Specifications 1
1-9
General Specifications
Power
Voltage
100 V Setting ...................................................... 90 V to 110 V
120 V Setting ...................................................... 108 V to 132 V
220 V Setting ...................................................... 198 V to 242 V
240 V Setting ...................................................... 216 V to 264 V
Frequency ............................................................... 47 Hz to 440 Hz. Automatically sensed at power-on.
Power Consumption................................................ 28 VA peak (12 Watt average)
Dimensions
Height...................................................................... 88 mm (3.46 in.)
Width....................................................................... 217 mm (8.56 in.)
Depth ...................................................................... 297 mm (11.7 in.)
Weight..................................................................... 3.6 kg (8.0 Ibs.)
Shipping Weight...................................................... 5.0 kg (11.0 lbs)
Display
Vacuum Fluorescent Display, dot matrix
Environment
Temperature
Operating ............................................................ 0 °C to 55 °C
Storage ............................................................... -40 °C to 70 °C
Warm Up............................................................. 1 hour to full uncertainty specifications
Relative Humidity (non-condensing)
Operating ............................................................ 0 °C to 28 °C <90 %
28 °C to 40 °C <80 %
40 °C to 55 °C <50 %
Storage ............................................................... -40 °C to 70 °C <95 %
Altitude
Operating ............................................................ 2,000 Meters
Storage ............................................................... 12,000 Meters
Vibration and Shock................................................ Complies with Mil-T-28800E Type III, Class 5 (Sine Only)
Safety
Designed to comply with IEC 61010-1:2000-1, UL 61010-1A1, CAN/CSA-C22.2 No. 61010.1, CAT I 1000V/CAT II 600V
EMC
Designed to comply with IEC 61326-1:2000-11 (EMC) when used with shielded communications cables. This Meter has
shown susceptibity to radiated frequencies greater than 1 V/m from 250 to 450 MHz while in the 100 µA range.
Triggering
Samples per Trigger ........................................... 1 to 50,000
Trigger Delay ...................................................... 0 S to 3600 S; in 10 µS increments
External Trigger Delay ........................................ <1 mS
External Trigger Jitter ......................................... <500 µS
Trigger Input ....................................................... TTL Levels
Trigger Output..................................................... 5 V maximum. (Open collector)