SDS1000X HD Series User Manual
82 i n t.s i g lent . c om
To set window trigger via the Relative window type
When the window trigger type is set to "Relative",
touch the trigger descriptor box. The pop-up menu will
show two user-defined parameters: Level +/-Delta"
and "Level Center".
The above two parameters can be set in the following two ways:
1. Select the parameter in the Level +/-Delta area of the quick menu, then set the parameter
value by the virtual keypad or the Level knob on the front panel. To set the center level is similar.
2. Directly use the Level knob on the front panel. Press the knob to switch between “Level +/-Delta"
and "Level Center", and rotate it to set values.
Note: “Level +/-Delta” represents half of the actual window area. For example, when the value is
200mV, it represents a range of ±200 mV, which is a 400mV window.
Holdoff, coupling, and noise reject can be set in window trigger, see the sections "Holdoff", "Trigger
Coupling" and "Noise Reject" for details.
15.5.7 Interval Trigger
Trigger when the time difference between the neighboring rising or falling edges meets the time limit
condition.
When the trigger condition is set as an interval between two neighboring rising edges and it is less
than the set time value, the trigger diagram is as follows: