6.Specification
70
6. Specification
This chapter lists the technical specifications and general technical
specifications of the spectrum analyzer. Unless otherwise stated,
the technical specifications apply to the following conditions:
The instrument has been preheated for 30 minutes before use.
The instrument is in the calibration cycle and has been
self-calibrated.
"Typical" and "nominal" for this product are defined as follows
Typical: Refers to the performance of the product under certain
conditions.
Nominal: Refers to the approximate value under product
application process.
Frequency
Frequency Range
XSA1015P
TG
9.000 kHz to 1.500000000
GHz
XSA1032P
TG
9.000 kHz to 3.200000000
GHz
XSA1036P
TG
9.000 kHz to 3.600000000
GHz
XSA1075P
TG
9.000 kHz to 7.500000000
GHz
Frequency Resolution
1 Hz
Internal Reference Frequency
Reference Frequency
10 MHz
Reference Frequency
Accuracy
±[ (days since last calibrate × freq aging rate)
+ temperature stability + initial accuracy ]
Initial calibration
accuracy
<1 ppm
Temperature stability
0 to 50, reference to 25
< 0.5 ppm
Aging rate
<1 ppm/year
Frequency Readout Accuracy