4.User Notice
15
receiver through an N type male connector, users can purchase
this option if required.
!
CAUTION
Input voltage at RF input port must not be higher than 50 V
DC to avoid damage to the attenuator and input mixer
tracking generator.
3. RF Input 50Ω
The RF input may be connected to a device via a N type connector
!
CAUTION
When input attenuator is higher than 10 dB, the RF port
input signal must be less than +30 dBm.
4.5Rear Panel
9 7 6 5 4 3
1
2
8
Figure 4-6 Rear Panel
Table 4-4 Rear Panel Description
NO.
Name
Description
Handle
Stow the handle for mobile use.
AC power
connector
AC: frequency 50H10, single-phase
alternative 220V±15 or 110V±15
Stool
To adjust the angle of the device