5.Menu Interpretation
31
order to sweep the adjacent channels manually.
You can modify this parameter using the numeric keys, knob,
direction keys
soft keyboard interface, or dragging trace
.
Frequency step lends itself to detect the harmonic waves and
bandwidths that are beyond the current span.
For example, for order of harmonic of a 300 MHz signal, you can
use set both the center frequency and frequency step to 300
MHz, and press the up direction key continuously to increase the
center frequency to 600MHz, that is secondary harmonic. Press
frequency steps to increase center frequency by 300MHz, which
reaches 900MHz. [CF Step Auto Man] shows the auto or manual
mode to setting the steps. When step is under manual mode,
press [CF Step Auto Man] to return to auto mode.
5.1.5 [Freq Offset]
You can set a frequency offset to displayed frequency value,
including freq marker value. This movement won’t influence sweep
frequency range.
While this function activated (frequency offset isn’t 0), you can
modify this parameter using the numeric keys, knob or direction
keys.
5.1.6 [Freq Ref Int Ext]
Set the reference frequency as internal or external input, this is
regarded as whole device reference. If the external signal is not
locked according to judgment after switching to external, the prompt
“external reference not locked” will pop up and it will switch back to
internal automatically.
5.2 SPAN
Set the spectrum analyzer to span mode. When press SPAN,
[Span],[Full Span],[Zero Span] and [Last Span] will be available to
configure. You can modify span using the numeric keys, knob or
direction keys. Use numeric key or [Zero Span] to clear span.